6.1.4.
"Section 15.1.4 Saving the Parameter File" 
Run the Application Software for Service, and then save the parameter file for the original CONTROL Board
to the replaced CONTROL Board.
(1) Run "LTabMonM.exe" within the folder 
      "UB-T580_ServiceSoft".
(2) Click the "Cancel" in the "Save As" dialog box.
(3) Click the "Load from file" in the "Sensor Adjust" dialog 
   
box.
(4) Select the parameter file obtained, and then click the "Open".
(5) Click the "OK" in the "LTabMonM" dialog box.
(6) Click the "Write to flash" in the "Sensor Adjust" dialog box.
(7) After confirming the content in the "LTabMonM" dialog box,
      click "Yes".
(8) Confirm that the message shown on the figure is displayed.