4-8. RTC TEST
The self-test of RTC module can be done.
Procedure:
1.  Enter the test mode.
 
(Refer to “SETTING THE TEST MODE” on page 15)
2.  Press the [
>
] or [
.
] keys to select the “RTC TEST”, and 
press the [
N
] key.
3.  The self-test of RTC module is executed, and result is dis-
played.
 
(Displayed characters/values in the following fi gure are ex-
ample)
LCD Display
RTC self-test result
<RTC TEST>
OK
Releasing method:
Press the [STOP/
x
] key, return to test mode menu.
4-9. BATT A/D
A/D value (voltage value, temperature value and IC operation 
state) acquisition can be performed.
Procedure:
1.  Enter the test mode.
 
(Refer to “SETTING THE TEST MODE” on page 15)
2. Press 
the 
[
>
] or [
.
] keys to select the “BATT A/D”, and 
press the [
N
] key.
3.  A/D value (voltage value, temperature value and IC operation 
state) are acquired, and they are displayed.
 
(8 bit (The bit shift after LSB side 2 bit omission))
 
Displayed values are always updated.
 
(Displayed characters/values in the following fi gure are ex-
ample)
LCD Display
<BATT A/D>
BAT:A7  TEMP:72
XST: A2
Releasing method:
Press the [STOP/
x
] key, return to test mode menu.
4-10. LOW 
DET
LOWBATT_XDET (Battery voltage decrease detection signal) 
can be checked.
Procedure:
1.  Enter the test mode.
 
(Refer to “SETTING THE TEST MODE” on page 15)
2. Press 
the 
[
>
] or [
.
] keys to select the “LOW DET”, and 
press the [
N
] key.
3.  The state of LOW DET is displayed.
 
(Displayed characters/values in the following fi gure  are  ex-
ample)
LCD Display
<LOW DET>
LOW DET : OFF
 
OFF : LOWBATT_XDET=1 (H)
 
ON  : LOWBATT_XDET=0 (L)
Releasing method:
Press the [STOP/
x
] key, return to test mode menu.
4-11. VBUS
VBUS detection signal can be checked.
Procedure:
1.  Enter the test mode.
 
(Refer to “SETTING THE TEST MODE” on page 15)
2.  Press the [
>
] or [
.
] keys to select the “VBUS”, and press 
the [
N
] key.
3.  The state of VBUS_XDET is displayed.
 
(Displayed characters/values in the following fi gure  are  ex-
ample)
LCD Display
<VBUS>
OFF
 OFF 
VBUS_XDET=1 
(H)
 ON 
VBUS_XDET=0 
(L)
 
Key input settings under this mode of operation:
 
[
N
] : It shifts to the suspension state.
 
Press the [RESET] button, in order to be released from the 
suspension state.
Releasing method:
Press the [STOP/
x
] key, return to test mode menu.