Read Sharp MX-PEX1 (serv.man14) User Guide / Operation Manual online
C
OLOR
M
EASUREMENT
I
NSTRUMENTS
96
9 If you have not already done so, remove the shipping spacer from the measurement
page slot.
10 Insert the end of the Color Reflection Reference strip with the arrow into the calibration strip
entrance on the DTP41, centering it below the alignment mark (see the diagram on
page 91
).
Insert the strip about 3 inches, past the front idler rollers, until it rests against the rear
drive rollers.
drive rollers.
11 Click Read Reference.
The DTP41 pulls the calibration strip through automatically.
A message appears in the Status area when calibration is finished.
12 Click Done to close the Instrument Calibration dialog box.
13 Close the Utilities dialog box.
This completes DTP41 calibration.
Your computer prompts you when the DTP41 requires calibration. When the prompt
message appears, click Calibrate Now and follow the instructions, starting with
message appears, click Calibrate Now and follow the instructions, starting with
step 8
on
page 95
.
After calibrating the DTP41, calibrate the Fiery X3eTY2 (see
page 54
).
Using the ED-100
The ED-100 is a hand-held instrument that measures the density of printed ink and toner.
Like the DTP32, the ED-100 is a reflection
Like the DTP32, the ED-100 is a reflection
densitometer
that you can use to calibrate the
Fiery X3eTY2.
The ED-100 is available as a separately purchased option.
Setting up the ED-100
To use the ED-100 for calibration purposes, you must first set it up for use with your
computer. For complete instructions on setting up the ED-100, see the documentation that
accompanies the ED-100.
computer. For complete instructions on setting up the ED-100, see the documentation that
accompanies the ED-100.
The following illustration shows the components of the ED-100.
1
Indicator light
2
Measure button
3
Sample aperture
(underneath)
(underneath)
1
2
3