AL-2021. Service Manual - Sharp Copying Equipment Service Manual (repair manual). Page 26

Read Sharp AL-2021 Service Manual online

AL-2041  UNPACKING AND INSTALLATION  5 - 9
3) Remove the rear cover for the document glass.
<1> Remove the two screws and then remove the right glass
holder.
<2> Slide the rear cover for the document glass to remove it.
<3> Remove the table glass.
4) Attach the Scanner condensation prevention mylar at the 3
positions on the rear side of the main unit as described below.
Note: The hole should be covered with the mylar.
Align the edge of the mylar to the R part (the yellow line in the
diagram below) so that the hole of the metal plate is covered as
much as possible.
Align the edge of the mylar to the R part (the yellow line in the
diagram below) so that the hole of the metal plate is covered as
much as possible.
Attach along the edge of the projection (the yellow line in the
diagram below).
5) Attach the Optical right hole mylar B at the 2 positions shown in
the diagrams below which are at the top of the rear side of the
main unit.
Note: The holes should be covered with the mylar.
Attach along the edge of the cushion (the yellow line in the dia-
gram below).
Align with the inside line of the bent part (the yellow line in the
diagram below).
Align with the raised part (the yellow line in the diagram below).
Match the center of the mylar (in the horizontal direction) to the
center of  the raised part.
1
2
3
Stick the excessive part on the side.
Stick the excessive part on the side.
Page of 120
Display

Click on the first or last page to see other AL-2021 service manuals if exist.