AVR 445 (serv.man3) - Harman Kardon Audio EMC - CB Certificate (repair manual). Page 13

Read Harman Kardon AVR 445 (serv.man3) EMC - CB Certificate online

 
Harman Kardon Inc.                                                Page 12 of 211 
                                  AVR445 (Harman Kardon)
 
EMC Test Report 
Reference No.: NK2GE 642.EMS 
 
 
10. Electrostatic Discharges (EN 61000-4-2) 
 
10.1 Test criteria and test procedure 
 
      Operating environment; 
 Temperature: 
20  ℃ 
 
Relative Humidity: 
40 % 
 
 
A ground reference plane was located on the floor, and connected to earth via a low impedance connection 
The return cable of the ESD generator was connected to the reference plane. In case of a floor standing 
equipment, EUT was placed on the reference plane on 10cm of insulating support 
In case of table-top equipment, EUT was placed on a wooden table 0.8m above the reference planed On the 
floor. A horizontal coupling plane (HCP) of 1.6*0.8m was placed on the table, and connected to the reference 
plane via a cable with a 470 ㏀  resistor located in each end(0.5mm insulating support between EUT and HCP). 
In both cases a vertical coupling plane (VCP) of 0.5*0.5m was located 10cm from the EUT’s sides. 
The VCP was connected to the reference plane in the same matter as the HCP. 
The test was done by applying contact and air discharges to the EUT itself, and contact discharges to the 
HCP and VCP. When applying the discharges to the HCP, the tip of the generator was located at each side of the 
EUT, at 10cm distance. When applying the discharges to the VCP the tip of the generator was located at the 
middle edge of the VCP. The VCP was located 10cm from each side of the EUT. 
Contact discharges of  ±4kV were applied to various points of the EUT at conductive surfaces, and to the 
HCP/VCP. Air discharges of  ±8kV were applied to various points of the EUT at non-conductive surfaces. 
Ungrounded equipment, or ungrounded part(s) of equipment, cannot discharge itself similarly to class I 
mains-supplied equipment. If the charge is not removed before the next ESD pulse is applied, it is possible   
that the EUT or part(s) of the EUT be stressed up to twice the intended test voltage.   
Therefore, double-insulated equipment could be charged at an unrealistically high charge, by accumulating   
Several ESD discharges on the capacitance of the class II insulation, and then discharge at the breakdown   
voltage of the insulation with a much higher energy. In the case of class II & III equipment. 
To avoid of unrealistically high charge, it is possible to ionize “using ionizer” before each discharge. 
 
Instrumentation used during this session: 
 
Instruments 
Manufacturer 
Model 
Serial Number 
Due to Calibration
ESD Simulator 
Noiseken 
ESS-2000 
0199C02406 
2006. 12 
Discharge Gun 
Noiseken 
TCP-815P 
1199C02485 
2006. 12 
 
10.2 Test Parameters 
 Port: 
Enclosure 
 
Test Engineer: 
J. Y. Yu 
 Basic Standard:  EN 61000-4-2:1998+A2: 2001 
Performance Criteria:  B 
 Test Level: 
Air Discharge: 
±4, 8kV 
 
 
Contact Discharge: 
±2, 4kV 
 
Page of 127
Display

Click on the first or last page to see other AVR 445 (serv.man3) service manuals if exist.